The DVC-4 Digital Vibration Controller test setup and control parameters are saved in comma-delimited sequential files; the default file names are DvcFileS.* for Sine, DvcFileR.* for Random and DvcFileC.* for Classical Shock testing, where * is a user chosen 3-character extension such as a number from 001 to 999 or combination like B08, etc. Any text editor such as Windows Notepad can be used to open and review or modify the files.

 

The file consists of a 64 by 7 array plus 3 text strings at the end; each line of the array has 6 numbers followed by a comment text string. The numbers represent a Setup parameter, Frequency, Mode, Level, Limit, and Alarm. The Setup numbers reflect all of the menu selections and vary some between the three programs; a complete listing of them is given in the table below. The 3 text strings at the end are the test title, accelerometer model, and accelerometer serial number.

 

Revised 1/4/00

Setup #

DSC- Sine

DRC-Random

CSC-Shock

Format

0

Control mode

Same

-

1=manual,2=average,3=extremal

1

Lower limit- Hz

Test duration

Pulse shape

 

2

Upper limit- Hz

Pretest level- db

Pulse amplitude

 

3

Sweep type

Filter (window)

Duration-msec

 

4

Sweep rate

Clipping ratio

Pulse polarity

 

5

Total sweeps

Abort limits- db

Repetitive pulses

 

6

Sweep rate

Frequency range

Repetition rate

 

7

-

-

Total pulses

 

8

Abort limits- db

-

Same as Sine

+/- Db.

9

Ch#1 sensitivity

Same

Same

Mv/g

10

Ch#2 sensitivity

Same

Same (reserved)

Mv/g

11

Ch#1 current

Same

Same

1=Off, 2=On

12

Ch#2 current

Same

Same (reserved)

1=Off, 2=On

13

-

-

Sample rate

 

14

-

-

Equalized

 

15

Base address

Same

Same

 

16

Update rate

Same

Pretest level- Db.

 

17

View Display

System max Acc

-

 

18

System max Disp

Same

-

 

19

System max Vel

Same

System max Disp

 

20

Remote inputs

Same

System max Vel

 

21

System max Acc

-

Same as Sine

 

22

Units

Same

Same

1-English, 2=metric

23

Frequency range

-

Remote inputs

1=Active, 2=Off

24

-

-

View Disp.

1=Off, 2=On

25

-

-

View Velocity

1=Off, 2=On

26

-

-

View Acc.

1=Off, 2=On

27

-

-

Max drive- volts

 

28

-

-

-

 

29

-

-

-

 

30

-

-

-

 

31

Dwell point- F1

-

System gain

 

32

Dwell point- T1

-

-

 

33

Auto-print

Same

Same

0=Off, 1=On

34

Auto-save

Same

Same

0=Off, 1=On

35

Dwell point- F2

Spectral Check

-

0=Off, 1=On

36

Dwell point- T2

-

-

 

37

Low gain trip

Same

Same

Db.

38

Open loop trip

Same

Same

Db.

39

Ch#3 sensitivity

Same

Same (reserved)

Mv/g

40

Ch#4 sensitivity

Same

Same (reserved)

Mv/g

41

Ch#3 current

Same

Same (reserved)

1=Off, 2=On

42

Ch#4 current

Same

Same (reserved)

1=Off, 2=On

43

Ch#1 control

Same

-

1=Off, 2=On

44

Ch#2 control

Same

-

1=Off, 2=On

45

Ch#3 control

Same

-

1=Off, 2=On

46

Ch#4 control

Same

-

1=Off, 2=On

47

Ch#1 view

Same

-

1=Off, 2=On

48

Ch#2 view

Same

-

1=Off, 2=On

49

Ch#3 view

Same

-

1=Off, 2=On

50

Ch#4 view

Same

-

1=Off, 2=On

51

Zoom-X axis

Same

Same

1=In, 2=medium, 3=out

52

Zoom-Y axis

Same

Same

1=In, 2=medium, 3=out

53

Plot color

Same

Same

0=Black, 1-white

54

-

-

Load comp.

0=minimum to 4=maximum load

55

 

 

 

 

56

 

 

 

 

57

 

 

 

 

58

 

 

 

 

59

 

 

 

 

60

 

 

 

 

61

 

 

 

 

62

 

 

 

 

63

 

 

 

 

 

 

wave